Skip to content

ORT-ASC-950-DP-G Automatic Sample Changer

ORTEC Model ASC-950-DP Alpha Beta Radiation Counting Automatic Sample Changer
  • Automatic Sample Changer
  • 2 in. (5.1 cm) Standard Planchet Diameter
  • Standard 50 Sample Capacity (Optional 100 sample)
  • Gross Alpha/Beta Counter
  • Dual Phosphor Scintillator Detector
  • Gasless
  • Lightweight
  • Touch Screen Interface
  • Easy Window Replacement
  • Radon Compensation For Air Filters
  • Optional Cosmic Guard Detector
  • Optional Lead Shielding to Reduce Background Levels
  • Literature +

  • More Information +


    The ORT-ASC-950-DP-G is a high-speed counter equipped with a dual phosphor scintillator type detector that supports gasless operation and easy portability. Cosmic background radiation reduction is possible with the optional Anti-Coincidence Guard Detector. DUO Express software allows for easy system programming, data management, and control chart maintenance. Multiple instruments can be connected for user convenience. This sample counter is ideal for standard Health Physics applications such as smears and air filters.

    • A built-in color touch screen interface
    • Ability to export data directly to a computer using the PIC Link option
    • Ability to program routines and QC charts on a PC running the included DUO Express software, then transfer the routines to the instrument using a direct-connect PIC Link cable
  • Specifications +


    Optimized for Higher Energy Betas
    Typical Efficiency
    Source Top 1/8" Bottom 1/4" Bottom 5/16" Bottom
    Pu-239 34% 30% 24% 21%
    SrY-90 45% 40% 35% 34%
    Tc-99 24% 21% 19% 18%
    Background
    Condition Warranted (CPM) Typical (CPM)
    Alpha Beta Alpha Beta
    With Lead and Guard <0.2 <25 0.05 12 
    With Guard, No Lead <0.2 <30 0.03 15
    With Lead, No Guard <0.2 <55 0.08 35
    No Guard or Lead <0.2 <65 0.06 40

    Optimized for Lower Energy Betas
    Typical Efficiency
    Source Top 1/8" Bottom 1/4" Bottom 5/16" Bottom
    Pu-239 35% 30% 24% 22%
    SrY-90 47% 41% 36% 35%
    Tc-99 26% 24% 20% 19%
    Background
    Condition Warranted (CPM) Typical (CPM)
    Alpha Beta Alpha Beta
    With Lead and Guard <0.2 <25 0.05 13
    With Guard, No Lead <0.2 <30 0.05 23
    With Lead, No Guard <0.2 <55 0.05 41
    No Guard or Lead <0.2 <65 0.06 55

    Instrument Specifications
    Sample Detector 2 in (5.1 cm) dia. ZnS + plastic dual phosphor
    Window Aluminized, 80 µg/cm2
    Sample Size 2 in (5.1 cm) diameter planchets, 1/8, 1/4, or 5/16 in (3.2, 6.3, or 7.9 mm) deep 
    Sample Transport 50 sample automatic changer with barcode reader and sample range 1–999
    Count Modes Alpha only, Alpha/Beta simultaneously, Alpha then Beta
    Interface Touch Screen
    Data Export USB PIC Link, USB Flash Drive
    Power 110–240 V ac, 50–60 Hz Input
    Dimensions 50 Sample: 36 in H x 23.5 in W x 23.5 in D (91.4 x 59.7 x 59.7 cm)
    100 Sample: 54 in H x 23.5 in W x 23.5 in D (137.2 x 59.7 x 59.7 cm)
    Weight 50 Sample: 110 lb. (49.9 kg)
    100 Sample: 120 lb. (54.4 kg)
  • Ordering Information +


    Model Description
    ORT-ASC-950-DP-G Automatic Sample Changer with Cosmic Guard. Includes 1 ea. PIC Link Cable, 1 each Standard Carrier Set, and 55 ea. 1/8 in Inserts.
    Options
    PIC-110-0173 Cosmic Guard Detector, 5 lb (2.3 kg)
    ORT-ASC-950-SHLD Lead Ring, 50 lb (22.7 kg)
    PIC-ACC-02 Cart
    ORT-ASC-950-OPT-04 100 Sample Option (adds ~18 in. to height and 10 lbs.)
    ORT-ASC-950-OPT-05 50 Sample Stack, Stand, and Carrier Set
    ORT-ASC-950-OPT-06 100 Sample Stack, Stand, and Carrier Set
    PIC-140-0305-01 Carrier Inserts, 50 ea., 1/8 in.
    PIC-140-0305-02  Carrier Inserts, 50 ea., 1/4 in.
    PIC-700-0007 Planchet, 2 inch x 1/8 inch, set of 100
    PIC-700-0008 Planchet, 2 inch x 1/4 inch, set of 100
    PIC-700-0009 Planchet, 2 inch x 5/16 inch, set of 100
    ORT-VISTA FC
    Health Physics Applications Software