HPGe Radiation Detector Electronics

The front-end electronics include the input field-effect transistor (FET) and the feedback element of the charge-sensitive preamplifier. The feedback element comprises a capacitor and a resistor (passive feedback) or a transistor (active feedback). Preamplifiers with active feedback, such as the transistor-reset preamplifier (TRP), are used for achieving ultra-high count rates.

 

A temperature-sensing element, also located in the cryogenic vacuum enclosure, automatically shuts off the HV power supply if the detector temperature begins to rise. The front-end electronics are enclosed in the cryogenically-cooled enclosure to minimize electronic noise. Both the input FET and the feed-back element have optimum noise characteristics at ~120 K.

 

Image: Simplified Electronic Diagram Showing the Detector and the "Front End" of the Preamplifier in a Cryogenically Cooled Germanium Radiation Detector. The feedback element may be either a resistor or instead an electronic switch that resets the FET gate voltage to zero whenever a given threshold voltage is reached.

ORTEC Detector Front End Electronics