HPGe Radiation Detector Types and How to Choose
Categories of High Purity Germanium (HPGe) Radiation Detectors
ORTEC offers a wide range of HPGe radiation detector solutions which cover low energies for X-rays up to gamma rays in the 10 MeV and above range. These detectors are manufactured using high purity germanium (HPGe) of both P- and N-type. All detectors in these two classes are cryogenically cooled, using either liquid nitrogen or electro-mechanical coolers.
How to Choose the Right HPGe Radiation Detector for Your Application
Choosing the correct detector for your specific application involves many criteria. No single table or graph can uniquely specify a detector for a particular measurement. The energy range chart above, and the key attribute table below should provide an excellent start in making the right choice.
| Detector Series | Typical Counting Geometry | Relative Efficiency Range (%) | FWHM Resolution (in keV) @ Specified Energy | ||
|---|---|---|---|---|---|
| 5.9 keV | 122 keV | 1332 keV | |||
| GEM P-type coaxial |
Point source Large surface area source Marinelli beaker |
10-150 | n/a | 0.80-1.30 | 1.8-2.3 |
| Profile C P-type coaxial |
Point source Large surface area source Marinelli beaker |
20-175 | 0.60-1.10 | 0.8-1.30 | 1.8-2.3 |
| Profile M P-type coaxial |
Point source Large surface area source |
20-175 | n/a | 0.88-1.30 | 1.8-2.3 |
| Profile S P-type semi-planar |
Point source Large surface area source |
7-65 | 0.40-0.50 | 0.65-0.70 | 1.8-1.9 |
| Profile SP P-type semi-planar |
Point source Large surface area source |
7-65 | 0.30-0.43 | 0.59-0.63 | 1.8-1.9 |
| GMX N-type coaxial |
Point source Large surface area source Marinelli beaker Neutron background |
10-100 | 0.60-1.20 | 0.85-1.40 | 1.8-2.5 |
| GLP P-type planar |
Point source Small surface area source Timing |
n/a | 0.19-0.38 | 0.51-0.63 | n/a |
| GWL P-type reverse coaxial (well) |
Small volume source inside of the well | 90-450 cc active volume (15%-100% rel. eff.) |
n/a | 1.2-1.4 | 2.1-2.3 |