Skip to content

Compton Suppression Counting Systems

Compton-Suppression Systems are used to reduce the background continuum for low-background counting, thus improving the Minimum Detectable Activity and overall spectrum quality – particularly for small volume samples such as air filters and petri dishes.

ORTEC has supplied dozens of Compton-Suppression Systems throughout the world to meet the demanding needs of very low background counting applications.

ORTEC has evaluated a large variety of HPGe detectors with a range of performance parameters in efficiency, resolution, and Peak-to-Compton values to determine the optimal combinations of instrumentation to achieve the best overall performance.
  • Literature +

  • More Information +


    Basics of Compton-Suppression Counting
    In a typical low-background system, great effort is made to reduce the inherent radioactivity in the counting system components: the detector, lead shield, and air inside the shield compartment. These low-background systems tend to focus on reducing only cosmic and natural sources of background in germanium spectroscopy systems.

    Compton-suppression systems are designed to reduce the background levels observed in these typical counting systems. While low-background systems remove most of the constituents which add peaks to the spectrum, they do not address the main contributor to the background continuum: Compton-Scattering Events. Compton scattering occurs when the full energy of an incident photon is not completely absorbed by the HPGe detector and thus exits the detector leaving only part of its energy to be counted.  This partial energy peak appears in a gamma-ray spectrum as a random event below the full energy peak in what is referred to as the Compton Continuum.

    The ratio of the full-energy peak to the Compton Continuum is called the Peak-to-Compton (or P/C) ratio. In a standard HPGe detector, it is common to have a Peak-to-Compton ratio between 40:1 and 60:1 for the 1.33 MeV peak of 60Co. Larger detectors can have a P/C ratio of nearly 100:1.

    Because the escaping energy is a photon, it is possible to collect that energy with another detector. This is typically done with a large crystal made of a less expensive material such as NaI, and is known as a shield detector. By correlating events in the HPGe and the shield detector with timing electronics, events counted in the shield detector can be used to reject simultaneous events in the HPGe detector. The result is a suppression of the Compton continuum. In a Compton-suppression system, Peak-to-Compton ratios in excess of 1300:1 are achieved with a 60% relative efficiency N-type detector. This results in a reduction in background of approximately a factor of 10 and in MDA by a factor of more than 3.

    Factors in Choosing HPGe Detectors for Compton Suppression
    The effectiveness of a Compton-suppression system depends on the ability (efficiency) of collecting the scattered events from the HPGe detector in the shield detector. Because a photon has a probability of interaction with every material it encounters, it is necessary that there be as little material as possible between the active volume of the HPGe crystal and that of the shield crystal. Materials of concern include the following:

      What Characteristics to Choose  What Characteristics NOT to Choose
    HPGe Outer Contact Thin outer contact: use an ORTEC Gamma-X (GMX) detector which uses N-type germanium with an outer contact of 0.3 microns of boron.
    Standard P-Type detectors:  P-type detectors, like the GEM and Profile GEM-M, have a thick outer contact (~600 microns of Lithium). This contact is three orders of magnitude larger than a Gamma-X and the probability of a gamma ray being stopped in this contact (and lost from the suppression) increases substantially. 
    Extended range P-type detectors:  extended range P-type detectors, like the Profile GEM-C, only have a thin contact on front. This means that the majority of the crystal is surrounded with a thick lithium contact and should not be used for the same reason as a standard P-type.
    HPGe Crystal Cup Low density cup: ORTEC uses a 0.5 mm thick, low-background aluminum cup in fabricating its standard detectors.
    Copper crystal cups:  The use of copper as a crystal cup should be avoided because copper has a higher density and higher mass absorption coefficient. This increases the probability that a photon will not enter the shield detector.
    HPGe Endcap Low density endcap: Endcaps of 1.5 mm thick Carbon Fiber or low-background aluminum with a carbon fiber window are used in standard detectors.
    Magnesium endcaps:  Magnesium has a higher mass absorption coefficient than aluminum and increases the probability that a photon will interact with the endcap and not enter the NaI annulus and should be avoided.
    Copper endcaps:  As with the copper cups, copper endcaps used in some environmental low-background detector systems should be avoided.
    Air or Other Materials between the HPGe endcap and shield housing
    HPGe crystal diameter as large as possible: In a typical 83 mm diameter endcap, a detector up to 70 mm diameter may be used. This equates to a Gamma-X detector having a relative efficiency of up to 70%. Custom CSS can be made with larger N-type detector endcaps and a larger inner diameter NaI annulus.
     
    Shield Detector Housing
    Endcap diameter to fit within the annulus of the NaI detector: NaI annulus can be custom fit to the specifications when ordered.
     

     
    Those familiar with low-background detectors will notice that magnesium endcaps, copper endcaps, and copper crystal cups are often used in low-background detectors. While conventional thinking would state that a Compton-Suppressed Low-Background detector would be the optimum solution for very low-background counting, this is not the case. By using these materials in detector construction, the overall effect is to reduce the capability of the Compton-Suppression System. The tradeoff can be significant.

    Other Important Components of the Compton Suppression System
    To capture gammas that are scattered from the HPGe detector into the NaI annulus or a plug detector in coincidence and to minimize the Compton continuum, it is very important to properly setup the electronics read-out chain. Historically, ORTEC has offered an analog readout chain using NIM based electronic components. As digital electronics developed, ORTEC introduced a digital dual-MCA version. The digital version considerably simplifies gating setup both in terms of time and complexity.

    The digital version of the CSS significantly decreases the space needed for electronics and allows for a more compact setup with fewer components, as a result there is no need for a large NIM rack. Without the need for the NIM rack, to further minimize space efficiency, ORTEC has replaced the box style shield with a cylindrical style shield that is 4-inch (10 cm) thick lead with an inner shield height 4-inches (10 cm) taller than a typical commercial shield. The cylindrical shield is half the weight and almost half the foot print of the analog version. Additionally, the design of the top sliding door of the cylindrical shield allows the guard detector to be mounted inside the top door. This novel innovation simplifies sample loading and unloading to a single step, saving time and minimizing damage that can occur to the guard detector during loading and unloading of the sample each time. Removing the guard detector from within the NaI annulus increases the sample space that can be used for measurements without sacrificing 4-π geometry and the Compton Suppression rejection performance.
    ORTEC Analog Compton Suppression SystemORTEC Digital Compton Suppression System
  • Specifications and Ordering Information +


    The table below summarizes the typical CSS option. 

    An optional cosmic-veto upgrade wtih a plastic scintillator is available. Contact the factory to discuss specifications and additional electronic components required.

    For customized HPGe detectors (larger GMX or semi-planar HPGe detectors), annulus (larger I.D. or alternative materials (e.g. BGO)), shielding, or table, contact the factory.

    Key Components Model CSS-D-60
    Detector Default detector is N-type HPGe (GMX) 60% relative efficiency with 3.25” (83 mm) maximum diameter endcap, low-background Al cup, and Carbon Fiber endcap. Other detector sizes are optional up to 60% relative efficiency and up to 3.25” diameter endcap.
    NaI Annulus and Guard Detector 9”x9” NaI annulus with 3.5” I.D., 10.75” O.D. and FWHM @662 keV ≤11%.
    6”x2” NaI plug “pancake” detector with FWHM @662 keV ≤7%.
    Maximum Sample Size 3.5” diameter by 3” high (up to 750 mL).
    Electronics DSPEC-502A Digital Dual MCA with superior timing and gating performance and less than 30 minutes setup time, and 556H Power Supply for NaI annulus.
    Shielding
    Cylindrical 4-inch (10 cm) thick reprocessed lead with Tn/Cu inner liner. 
    Shield cavity is 11” diameter by 18” high.
    Optional cosmic-veto upgrade with a plastic scintillator is available.
    Software
    GammaVision Gamma Spectroscopy Software included. Computer is typically not included, but can be provided on request.
    Guaranteed Performance
    P:C at 500 keV averaged background region
    1300:1 or greater (with default GMX60 detector). 
    Note: Smaller GMX detectors will have lower P:C performance.
    Background Rate
    0.5 cps or lower with rectangular or cylindrical shields.

    For a complete system including specially designed lead shield, timing electronics, suppression shield detector, and HPGe detector, contact the factory.