Skip to content

WPC-1150-GFW Automatic Sample Changer


  • 3.5 in. (8.9 cm) or 5.0 in. (12.7 cm) Diameter Pancake Gas Flow Detector
  • Cosmic Guard Detector
  • Low Background
  • Custom Planchet Sizes Available
  • 50 Sample Capacity
  • Touch Screen Interface
  • Counting Gas Loss Detection
  • Literature +

  • More Information +


    The WPC-1150-GFW‑3.5 and WPC-1150-GFW-5 low background automatics are gas-flow proportional counters equipped with either a 3.5 in. (8.9 cm) or 5.0 in. (12.7 cm) detector configuration. Each counter incorporates a built-in color touch screen interface and is targeted toward Health Physics and environmental applications. They are ideal for large diameter air filters used in high-volume air samplers.
  • Specifications +


    Performance
    Typical Background
      WPC-1150-3.5 WPC-1150-5
    Alpha 0.3 cpm 0.4 cpm (0.5 cpm warranted)
    Beta 1.5 cpm 4 cpm (5 cpm warranted)
    Typical Efficiency
    230Th 27%
    90Sr/90Y 63%
    99Tc 41%
    63Ni 20%
    Cross Talk
    Alpha-Beta <0.1%
    Beta-Alpha <0.1%

    Instrument Specifications
    Sample Detector WPC-1150-GFW-3.5: 3.5 in. (8.9 cm) dia., pancake style gas flow proportional
    WPC-1150-GFW-5: 5 in. (12.7 cm) dia., pancake style gas flow proportional
    Window Aluminized, 80 µg/cm2
    Guard Detector 1 large gas flow proportional
    Counting Gas P-10, 60 cc/min at 10 psi
    Gas-PRO Fail-safe P-10 monitoring system senses breaches in gas flow
    Shielding 4 in. (10 cm) thick lead
    Sample Size Custom planchet sizes available, accommodates planchets up to 5/16 in. (7.9 cm) deep
    WPC-1150-GFW-3.5: active detector area of 3.5 in. (8.9 cm)
    WPC-1150-GFW-5: active detector area of 5 in. (12.7 cm)
    Sample Transport 50 sample automatic changer with barcode reader with sample range 1–999
    Count Modes Alpha only, Beta only, Alpha/Beta simultaneously
    Interface Touch Screen
    Data Export USB flash drive, RS-485
    Power 117 V ac, < 0.2 A, 50/60 Hz. 230 V ac, < 1.0 A, 50/60 Hz
    Environment 10–40 °C (50–104 °F). 20%–90% humidity, non-condensing
    Dimensions 25 in W x 36 in. D (62.5 x 91.4 cm)
    Weight 1800 lb (816 kg)
  • Ordering Information +


    Model Description
    ORT-WPC-1150-GFW-3 Automatic Sample Changer with 3.5 in. (8.9 cm) dia., pancake style gas flow proportional detector
    ORT-WPC-1150-GFW-5 Automatic Sample Changer with 5 in. (12.7 cm) dia., pancake style gas flow proportional detector
    Options
    PIC-700-0007 Planchet, 2 inch x 1/8 inch, set of 100
    PIC-700-0008 Planchet, 2 inch x 1/4 inch, set of 100
    PIC-700-0009 Planchet, 2 inch x 5/16 inch, set of 100
    ORT-Vista FC Health Physics Applications Software
    ORT-Vista 2000 Radiochemistry Applications Software