556 and 556H High Voltage Power Supply

ORTEC 556 and 556H High Voltage Power Supply
  • For use with photomultiplier tubes, microchannel plates, proportional counters, and Geiger-Mueller tubes
  • 0 to ±3 kV
  • 0 to 10 mA
  • Digital meter reads output voltage or current
  • Overload and short-circuited protected
  • External control of output voltage
  • 115 or 230 V ac, 47 to 63 Hz input power

The ORTEC Models 556 and 556H High-Voltage Power Supplies provide the noise-free, well-regulated, very stable high voltage necessary for proper operation of photomultipliers, ionization chambers, semiconductor detectors, electron multipliers, and many other devices.
  • Literature +

  • More Information +


    The Model 556 is housed in a double-width NIM module (per DOE/ER-0457T). The Model 556H is a stand-alone instrument designed for bench-top operation. The low-noise output is continuously adjustable from ±10 to ±3000 V dc with 0 to 10 mA load current. Noise on the output is <15 mV peak-to-peak, thereby ensuring the highest performance in high-resolution, semiconductor or scintillation spectroscopy systems.

    The front-panel digital meter allows visual monitoring of either the output voltage or the output current. 

    The output voltage can be controlled from ±10 to ±3000 V by application of an external input voltage of 0 to ±6.9 V dc. This feature is desirable for control applications and is standard on all units.

    The input power for Models 556 and 556H is taken directly from the ac line, either 115 or 230 V ac, 47 to 63 Hz.

    Overload and short-circuit protection networks permit operation into short circuits without damage to the instrument.
  • Specifications +


    Performance
    Output
    Polarity
    Positive or negative, selected by switch on rear panel.
    Output
    Range
    50 to 3000 V; minimum usable voltage 10 V.
    Output
    Load
    Capacity
    0 to 10 mA.
    Regulation ≤0.0025% variation in output voltage for combined line and load variations within operating range at constant ambient temperature.
    Temperature
    Instability
    <±50 ppm/°C after 30-minute warmup; operating range 0 to 50°C.
    Long
    Term
    Drift
    <0.01%/hour and <0.03%/24-hour variation in output voltage at constant input line voltage, load, and ambient temperature after 30-minute warmup.
    Output
    Ripple
    <15 mV peak-to-peak, 20 Hz to 20 MHz.
    Overload Protection Internal circuitry protects against overloads including short circuits.
    Resettability Output voltage can be reset to within 0.1%.

    Controls
    Power
    Front-panel toggle switch energizes unit when power cord is  connected to appropriate source, and an adjacent red LED lamp indicates when power is applied.
    Output
    Voltage
    One 6-position switch, one 5-position switch, and one 10-turn precision potentiometer; output voltage is the sum of the 3 settings.
    Meter Front-panel toggle switch selects display of output voltage in V or load current in mA.
    Polarity Rear-panel switch selects either positive or negative output polarity.
    Control Rear-panel locking toggle switch selects the reference source for the output voltage.
    • Int  Selects the internal reference source; the front-panel controls select the output voltage.
    • Ext  Selects the external reference source; output voltage is proportional to magnitude of reference input.
    AC Voltage Rear-panel slide switch selects either 115 V or 230 V ac input voltage.

    Inputs

    AC Power
    103–129 V or 206–258 V, 47–63 Hz, 70 W nominal at full output power; supplied through international standard IEC power connector on rear panel. Fuse rating: 1.5 A, 250 V fuse for 115 V AC operation or 0.75 A, 250 V fuse for 230 V AC operation.
    External
    Control
    Full range of output voltage can be based on an external dc reference level furnished through a rear-panel BNC connector; control voltage range is 0 through ±6.9 V DC; control voltage polarity must be the same polarity as that selected by the rear-panel Polarity switch; this input protected against over-voltages >±7 V. Input impedance >45 kΩ.

    Outputs

    Regulated
    DC Output
    The adjusted and regulated voltage, with selected polarity, is furnished simultaneously to the two SHV connectors on the rear panel.

    Indicator
    Meter
    Front-panel LCD display indicates output voltage in kV ±10 V or load current in mA ±10 µA. Load current is sum of external load current and internal load current. Internal load resistance is ~5 MΩ.

    Electrical and Mechanical
    Power
    Requirements
    115 or 230 V AC, 47–63 Hz, 70 W nominally at maximum output load.
    Weights 556   Net 3.6 kg (8.0 lb).  Shipping 4.5 kg (10.0 lb).
    556H   Net 5.7 kg (12.6 lb).    Shipping 6.6 kg (14.6 lb).
    Dimensions 556  Standard double-width NIM module, 6.90 X 22.13 cm (2.70 X 8.714  in.).
    556H  11.43 X 22.35 X 29.21 cm (4.5 X 8.8 X 11.5 in.).
  • Ordering Information +


    Model Description
    556 High-Voltage Power Supply
    556H High-Voltage Power Supply

    Optional Cable Accessories
    C-34-12 RG-59A/U 75-Ω Cable with one SHV female plug and one MHV male plug, 12-ft. length.
    C-36-12 RG-59A/U 75-Ω Cable with two SHV female plugs, 12-ft. length.