Input A accepts an input signal with a width of 50 ns or more and regenerates an internal signal that will be used for coincidence comparisons. The Input A signal width is adjustable for a resolving time of 100 ns to 2 µs, and this range is available with a front-panel control.
The function of each input is selectable, and its signal can be used for coincidence or anticoincidence or can be disabled. This permits various combinations of input signal relations to be selected without adding or removing cables in the system.
Another feature that simplifies operating flexibility without changing any cables is a selectable number of inputs that are required to satisfy a coincidence. For example, if the selector shown is set at 2, an overlap between any two inputs that are selected for the coincidence function will cause an output to be generated. If any one or more inputs are selected for anticoincidence, all outputs are inhibited while such signals are present. Because any combination of input signal effects can be selected easily at the front panel, the Model 418A is a Universal Coincidence unit that can be adapted to any coincidence system arrangement.