576-VM Vacuum Manifold and Control

  • For convenient connection of as many as four Model 576A Dual Alpha Spectrometers to a single pump
  • Screw-type vacuum valves for fine adjustment of pumping speed
  • Adds redundant valve controls for use with SOLOIST™ to allow fine adjustment of pumping speed

ORTEC Model 576-VM Vacuum Manifold and Control has four valves and terminations for connecting as many as four Model 576A Dual Alpha Spectrometers or SOLOIST Alpha Spectrometers to a single pump, so that any one unit may be vented without venting the remainder. The Model 576-VM Vacuum Manifold also includes a thermocouple vacuum gauge to monitor vacuum.

SPECIFICATIONS

Performance Helium leak checked to <3 x 10-8 atm cm3/s.

Operating Range 1 to 5000 Millitorr.

Controls Four screw-type vacuum valves.

Inputs Four flexible lines with 1/4 in. Swagelok® connectors for connection directly to 576A or SOLOIST modules.

Outputs 1-1/8 in. copper tube with NW-25 ISOKF flange termination for connection to user-supplied vacuum pump (minimum pumping speed of 5.6 scfm). May be connected directly to Model ALPHA-PPS Portable Pump Station. Flange may be cut off with tubing cutter if connection by rubber vacuum hose is desired.

ELECTRICAL AND MECHANICAL

Power Required 115 V ac/60 Hz or 230 V ac/50 or 60 Hz

Weight
Net 3 kg (7 lb).
Shipping 9 kg (20 lb).

Dimensions 18 cm high x 48 cm wide x 53 cm deep (7 x 19 x 21 in.); 19 in. rack mount.

ORDERING INFORMATION
 

Model Description
576-VM Vacuum Manifold and Control for 115 V ac
576-VM-230 Vacuum Manifold and Control for 230 V ac