RCAP-2 Ion Recoil Contamination Avoidance Package
The Ion Recoil Contamination Avoidance Package (RCAP-2) is an accessory for ORTEC’s OCTĘTE™ Plus Alpha Spectrometer. It is intended for alpha spectroscopists quantifying radionuclides which may contaminate charged-particle (ULTRA™ or surface barrier) detectors by ion recoil. To avoid recoil contamination, one can reduce the momentum of the positively-charged recoils by increasing the vacuum chamber pressure and negatively biasing the sample holders, to repel the ions from the detector. The result is achieved at some minor cost in energy resolution. In an OCTĘTE Plus equipped with RCAP-2, all sample holders are negatively biased at –12 volts. The RCAP-2 can control the pressure in the range of 10–3 Torr to one atmosphere. The pressure control package consists of:
Note that pressures below 1 Torr in the vacuum chambers can also be read and displayed with the OCTĘTE™ Plus on the front-panel digital display. Ordering Information
|