| Home | Applications | Contact Us | ||
| Products | Service | Training | ||
|
|
Research Applications Timing Download PDF (418 KB)
Simultaneous Time and Pulse-Height Spectroscopy with Detectors Having a
Single Output
Time Spectrometry Beyond 10 µs: Time-of-Flight Mass Spectrometry and
LIDAR From the measured flight time, t, the mass of the ion can be computed as
To achieve sub-nanosecond digital resolution, substitute the Model 9327 1-GHz Amplifier and Timing Discriminator and the Model 9308 picosecond TIME ANALYZER for the VT120A Preamplifier and the MCS-PCI or 9353. See Figure 2(B) and Application Note AN52. Because there is a dead time associated with the processing of each detected event in the MCS-PCI, 9353 and the Model 9308, these products are limited to operating in the single-ion counting/timing mode in the TOF-MS application. When the input is busy processing one event, it cannot respond to additional events arriving during the dead time. This means that the probability of detecting an event in the dominant peak in the time-of-flight spectrum must be limited to less than 1% during any scan. Mathematical dead time corrections can raise this limit to 10%. But in either case, the allowable ion rates are very low. This results in >10% statistical uncertainties for the counts in the peak, when the data acquisition period is much shorter than 1 second.Ion rates can be increased by a factor of more than 100 by employing the FASTFLIGHT™ or FASTFLIGHT-2™ Digital Signal Averager, as illustrated in Figures 3(A) and 3(B). This results in a factor of 10 improvement of statistical precision and detection limits. FASTFLIGHT eliminates the event-processing dead time by using a flash ADC to sample and digitize the analog signal from the detector at 0.5, 1, or 2-ns intervals. Multiple ions arriving within the detector pulse width are fully counted, because the detector and flash ADC respond linearly to the number of ions in the pulse. For more information, see Multichannel Scalers, Multiple-Stop Time Spectrometers, or Application Notes AN53 and AN54. A site dedicated to Mass Spectrometry is http://.www.signalrecovery.com/ms Fig. 4(A) shows the application of the MCS-PCI or 9353 to a LIDAR system for studying the concentration of compounds as a function of altitude in the atmosphere. Laser pulses scattered by molecules in the atmosphere are detected by a photomultiplier tube. The round-trip flight time of the photons is measured by the MCS unit to determine the altitude at which the scattering took place. The counting rate of the detected photons can be used to measure the concentration of specific compounds. In practice, the laser and the photomultiplier tube are incorporated into a system of lenses, designed to limit the field of view and to guaranty overlap between the volume excited by the laser and that viewed by the photomultiplier tube. Typically, two parallel systems are used to measure the response at different wavelengths. This allows differential absorption corrections to be applied. The latter technique is called Differential Absorption Lidar (DIAL).For the best resolution in LIDAR choose the instrumentation in Figure 4(B). Timing with Scintillation DetectorsFigures 5, 6, and 7 are examples of high-resolution time spectrometry with scintillation detectors. Coincident pairs of gamma rays from the radioactive source are detected in opposite scintillation detectors. Fig. 5 represents a “fast/slow” timing system. “Fast” refers to the fact that the fast anode signals from the photomultiplier tubes are used to derive the timing information. The dynode outputs are integrated by the 113 Preamplifier, processed as “slow” pulses in the 460 Amplifiers, and fed to the 551 SCAs to select the desired range of pulse heights (energies) in each detector. Strobing the TAC by the 414A Fast Coincidence output ensures that only pulses within the selected energy windows will contribute to the recorded time spectrum. Fig. 6 shows the effect of varying the width of the energy windows on the measured time resolution. Fig. 7 illustrates a “fast/fast” timing system, wherein the fast anode signal is used for both functions: 1) to derive the timing information, and 2) to select the range of pulse amplitudes. The upper and lower level discriminators in the Model 583B Constant-Fraction Differential Discriminator select the range of pulse amplitudes, while the constant-fraction zero-crossing discriminator adds the timing information. The advantage of this fast/fast system is the capability of operating at higher counting rates than is possible with the fast/slow system in Fig. 9. Time Spectroscopy with Germanium DetectorsThe system in Fig. 8 can be used to measure the timing characteristics of Ge detectors. Since the time resolution of the fast plastic scintillation detector is small compared to that of the Ge detector, the peak recorded in the time spectrum is characteristic of the Ge detector. Tables 1 and 2 summarize the time resolutions obtained with Ge detectors over a wide range of detector sizes. For further details see the sections on Amplifiers, and Fast-Timing Discriminators.
Timing with Silicon Charged-Particle Detectors Figures 13 and 14 are examples of coincidence spectroscopy systems intended for studying radioisotopes that emit multiple quanta of radiation in a single decay. The model 551 Timing Single-Channel Analyzers in Fig. 13 provide the slow timing information for determining the gamma rays striking the two detectors are truly coincident. The Timing SCA for Detector A is operated with a wide-open window to allow measurement of the entire energy spectrum for Detector A on the MCB. For Detector B, the window on the Timing SCA is adjusted to select a single gamma-ray energy. Consequently, the MCB records the energy spectrum from Detector A for all the gamma rays that are in coincidence with the gamma ray selected from Detector B. The system in Fig. 14 offers more powerful data acquisition capability than the scheme in Fig. 13. With Option 1, a true two-parameter data acquisition yields a three dimensional of the coincidence spectra from the two detectors. In addition coincidence-gated singles spectra from each detector can be recorded simultaneously via Option 2. Multi-parameter data acquisition and display is supported by the KmaxTMSoftware on either a Macintosh or IBM-compatible personal computer. In Fig. 14 a model 567 TAC/SCA is used to set the fast coincidence resolving time. This is more convenient than the scheme in Fig. 13 because the TAC time spectrum can be displayed via an ADC that is gated by the TAC/SCA while the SCA window is adjusted to accept only the true coincidence peak. The optional start gate on the TAC/SCA can be used to reduce dead time in the TAC.
Selecting the Type of Radiation by Pulse-Shape Analysis It is difficult to design a detector in which fast neutrons interact to produce a signal that yields good energy resolution. Consequently, neutron energies are normally determined by measuring the flight time of the neutron over a fixed distance. If the distance is s and the flight time is t, the energy of the neutron can be calculated as
Conversely, Equation (2) can be used to determine the mass of an unidentified particle, if the energy E is controlled. The method can be used to identify charged particles in nuclear reactions, or the molecular species in a time-of-flight mass spectrometer (see Figures 2 and 3). Fig. 17 shows a typical neutron time-of-flight spectrometer. Deuterium ions (21H+) are boosted to an energy of about 200-keV in an electrostatic accelerator, and directed to a target containing Tritium (31H). The resulting nuclear reaction, 31H (21H+, n) 42He, produces neutrons having an energy of 14.2 MeV, and recoiling alpha particles (42He) with an energy of 3.6 MeV. After scattering from the sample, the neutrons exhibit discretely different energies depending on (a) the nuclear states excited in the sample by inelastic scattering, and (b) the scattering angle.The time-of-flight spectrum is measured by the time interval between the alpha particle arriving at the ULTRA™ detector and the neutron arriving at the neutron detector. The high counting rate signals from the alpha-particle detector are delayed and used as the stop pulse, while the lower counting rate signals from the neutron detector are fed to the start input of the TAC. This reversed start/stop scheme reduces dead time in the TAC. A pulse shape analyzer, as described in Fig. 15, is used to reject gamma-ray background. The Model 552 SCA also serves to define the lower pulse-height threshold for accepting neutron signals. This threshold is critical in determining the detection efficiency of the neutron detector. Figure 1. Simultaneous Time and Pulse-Height Measurement with a Microchannel Plate Detector.
Calibration: 2.070
ps/channel
Figure 11. Typical Time Resolution vs. Detector Capacitance.
|
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||