C Series Annular Partially Depleted Silicon Surface Barrier Radiation Detector

Main Application
Backscattering from a collimated source or beam target; angular correlation measurements.
  • Literature +

  • Mounting Arrangements +


    A  This is a "ring mount"; i.e., the silicon wafer is offered on its ring without output connectors. This infrequently used arrangement is available on special request.
    T
     Microdot on the side of the can; without adjustable screws. (Open back.)

    Dimensions are given in millimeters.


    Detector Size (mm2)

    W (Nominal)
    Type A  Type T Transmission Mount
    X Y  X Y Z
     050  8.0 15.2 3.7
    19.4 7.9 9.9
     150  13.8 22.0 3.7
    26.1 7.9 9.9
     300  19.5 27.1 3.7
    31.6 7.9 9.9
     450 23.9 30.5 3.7
    34.8 7.9 9.9
     Tolerance ±0.5 ±0.3
    ±0.3
    ±0.3
    ±0.3
    ±0.3
  • Ordering Information +


    Supplied in T Mount unless specified otherwise by appropriate letter prefix; 4-mm hole is standard, 6- or 8-mm hole or other areas and depths available on special order.

    **First three digits of Model No. indicate total system resolution FWHM for 241Am, 5.486-MeV alphas, using standard ORTEC electronics and 0.5-µs shaping time constants. Beta resolution approximated by pulse generator width FWHM.


    Active Area (mm2)
    Guaranteed Maximum Resolution** (keV) Minimum Depletion Depth 100 µm
    Minimum Depletion Depth 300 µm  Minimum Depletion Depth 500 µm 
    α β Model No. Model No.  Model No. 
    50 17 9 C-017-050-100 C-017-050-300 C-017-050-500
    150 19 12 C-019-150-100 C-019-150-300 C-019-150-500
    300 21 14 C-021-300-100 C-021-300-300 C-021-300-500
    450 25 20 C-025-450-100 C-025-450-300 C-025-450-500


    Active Area (mm2)
    Guaranteed Maximum Resolution** (keV) Minimum Depletion Depth 1000 µm
    α β Model No.
    50 18 10 C-018-050-1000
    150 20 12 C-020-150-1000
    300 22 15 C-022-300-1000
    450 25 20 C-025-450-1000