Adding Delays Because of the
nature of the TAC circuitry, it is difficult to measure time
intervals <10 ns with good linearity. However, many measurements
involve start and stop signals that arrive within ±10 ns of each
other. The solution for these situations is to insert an appropriate
delay in the stop signal path. Selecting a delay in the range of 10
to 30 ns on an ORTEC Model 425A
Nanosecond Delay is usually sufficient to move the timing peak
into the linear region of the time spectrum. The stop delay can also
be adjusted to center the features of interest in the time spectrum. |