660 Dual 5-kV Detector Bias Supply
  • Two independent bias supplies in a single module for germanium and silicon detectors
  • 0–5 kV or 0–500 V at 0–100 µA
  • Remote shutdown feature compatible with ORTEC and TTL outputs from warmup sensors on germanium detectors
  • Reset safety feature on remote shutdown minimizes risk of preamplifier FET damage
  • Selected output polarity indicated before bias voltage is turned on
  • Automatic overload protection and overload indicator
The ORTEC Model 660 Dual 5-kV Detector Bias Supply contains two independently adjustable power supplies for furnishing the bias voltage to germanium detectors, silicon detectors, or ionization chambers. It can be used with any detector that draws less than 100 µA of current, and whose gain is insensitive to the applied voltage. The output voltages are continuously adjustable from zero to full scale with calibrated and locking 5-turn dials. Each supply provides two outputs controlled by the same dial: a 0 to 5 kV output, and a 0 to 500 V output. Two 10-segment bar-graph indicators verify that the selected voltages are being supplied at the outputs.

Security against accidentally changing the output polarity to the wrong state is ensured by two features. The selected output polarity for each supply is indicated by front-panel LEDs whenever the NIM bin power is turned on. Thus, the correct polarity can be verified before the HV ON/OFF switch is used to turn on the bias voltage to the detector. In addition, the side panel must be removed in order to alter the output polarity. This discourages unintentional changes.

The Model 660 includes a remote shutdown feature to protect the preamplifier FET against damage when a cooled germanium or Si(Li) detector warms up. Each supply includes a BIAS SHUTDOWN input that is compatible with the standard warmup sensor output on ORTEC preamplifiers. When the preamplifier signals a warmup condition, the Model 660 shuts off the bias voltage to that detector, and turns on a SHUTDOWN indicator light. The bias voltage remains off, independent of the signal from the preamplifier warmup sensor, until the shutdown mode is manually cancelled by pressing the RESET push button. This protects the preamplifier FET if the detector is cooling down with the HV ON/OFF switch accidentally left on. For further protection against operator error in the ORTEC shutdown mode, the bias shut-down input interprets a disconnected cable or a shorted cable as a warm detector, and responds by turning off the bias voltage. Some detector manufacturers provide a TTL logic level output from their detector warmup sensor. A board-mounted jumper in the Model 660 can be moved to the TTL position to make the bias shutdown input compatible with detectors supplying a TTL output. It is also possible to disable the bias shutdown feature by moving the board jumper to the BYPASS position. The Model 660 is shipped from the factory with both supplies set to the ORTEC mode.

The high voltage outputs are protected against overload. When the bias supply senses an excessive output current demand, it turns on the overload light and reduces the output voltage until the output current is within tolerable limits. Recovery from overload is automatic when the excessive current demand is eliminated.

ORDERING INFORMATION
 

Model    Description
660   5-kV Detector Bias Supply
C-24-12   RG-62A/U 93- Cable with two BNC male plugs; 12-ft length
C-36-12   RG-59A/U 75- Cable with two SHV female plugs; 12-ft length