428 Detector Bias Supply

PERFORMANCE

NOISE AND RIPPLE <0.0002%.

TEMPERATURE INSTABILITY £±0.02%/°C, 0 to 50°C.

LINE INSTABILITY Directly proportional to dc power supply instability (<±0.02% for 105 to 125 V ac when using one of the ORTEC Model 4002 Series Power Supplies).

CONTROLS AND INDICATORS

A/B (0–1000 V) Front-panel 10-turn direct-readout potentiometers for bias control.

POS/OFF/NEG Front-panel switch selects positive or negative outputs for both detectors.

CURRENT MONITOR Front-panel jacks for accommodating external meter in each output circuit.

OUTPUTS

A/B SHV connectors on front panel provide short-circuit-proof outputs for each detector; range 0–1000 V; positive or negative polarity for both detectors; impedance ~1.3 MW.

ELECTRICAL AND MECHANICAL

POWER REQUIRED +24 V, 165 mA; –24 V, 165 mA.

WEIGHT
Net
1.82 kg (4.0 lb).
Shipping
3.3 kg (7.25 lb).

DIMENSIONS NIM-standard double-width module 6.90 X 22.13 cm (2.70 X 8.714 in.) per DOE/ER-0457T.

ORDERING INFORMATION

Model   Description
428   Detector Bias Supply
C-36-12   RG-59A/U 75- Cable with two SHV female plugs, 12-ft length