ORTEC Logo

 
Home | Applications | Contact Us
Products | Service | Training

Specifications

Download PDF

Order Online

4002A 96-W Power Supply for NIM Bins
  • 4002a-rp.jpg (16325 bytes)Regulated dc power of ±12 V @ 2 A, and ±24 V @ 1.0 A; 117 V ac @ 0.4 A
  • 96 W of dc output up to 50°C
  • Operates from 100, 120, 220, or 240 V ac at 47 to 63 Hz
  • Attaches to 4001A or 4001C NIM Bins
  • Short-circuit and overload protected
4002a-fp.jpg (11589 bytes)The ORTEC Model 4002A Power Supply is designed to supply dc power to a NIM bin when the application requires ±12 V and ±24 V power. The 4002A Power Supply can be purchased separately for use with existing NIM bins, or it can be ordered attached to either a 4001A NIM Bin or a 4001C NIM Bin. The Model 4002A Power Supply is compatible with all ORTEC NIM bins, and with most standard NIM bins from other manufacturers. The 4002A Power Supply is designed to exceed recommended power supply specifications for Type 1 Class A supplies as defined in DOE/ER-0457T.

Regulated dc power supplied to the attached bin by the 4002A is rated at +12 V @ 2 A,
–12 V @ 2 A, +24 V @ 1 A, and –24 V @ 1 A. The maximum output power is 96 W at ambient temperatures up to 50°C. In addition, 117 V ac is available up to 0.43 A.

Protection against overload is provided in several ways. When the heat sink temperature exceeds 82°C, the red warning indicator is illuminated on the attached bin control panel. When the heat sink temperature exceeds 95°C, the power supply is automatically shut down, causing both the power and temperature indicator lights to turn off. Recovery from thermal overload is automatic when the thermal load is reduced. Output currents from the dc supplies are internally limited to 120% of their rated values by foldback circuits. This provides overload and short-circuit protection. Fuses protect the ac inputs to the power supply.

Designed for international use, the 4002A Power Supply can accept input voltages of 100, 120, 220, or 240 V ac at 47–63 Hz. A connector block at the rear of the 4002A is used to select the intended voltage range. The connector block also functions as a fuse holder and power cord connector. An international standard IEC power connector (CEE-22*) permits power cords and plugs that meet local electrical standards to be used for the input power. Control of the primary power is provided by the On/Off switch on the NIM bin control panel.

Connection of power and control lines to the NIM bin is provided by the standard interface connector specified in DOE/ER-0457T. Mechanical mounting of the power supply to the bin is with bolts utilizing the standard bolt pattern specified in DOE/ER-0457T.

*International Commission on Rules for the Approval of Electrical Equipment, standard number 22. 

ORDERING INFORMATION

The Model 4002A Power Supply may be ordered separately or assembled with a NIM bin, using the model numbers shown below.
 
Model   Description
4001A/4002A    NIM Bin and 96-W Power Supply (with ±12/24 V)
4001C/4002A   NIM Bin (with copper bus bars) and 96-W Power Supply
(with ±12/24 V)
4002A    96-WPower Supply (with ±12/24 V)