Single-Channel Pulse-Height Analyzers (SCA) Introduction
Download PDF
The amplitude of the analog pulse at the output of a spectroscopy
amplifier is typically proportional to the charge released in the
detector or to the energy of the detected event. Selection of a
range of signal levels at the output of the amplifier is equivalent
to the selection of a range of energies or charge for these events.
This selection can be accomplished by the use of discriminators and
single-channel analyzers (SCAs). A discriminator produces an output
logic pulse only if its input signal exceeds a preset threshold
level. A single-channel analyzer produces an output logic pulse only
if the peak amplitude of its input signal falls within the
pulse-height window that is established with two preset threshold
levels.
Figure 1 shows three pulses that might be provided from a main
amplifier to an integral discriminator. The first pulse has an
amplitude less than the adjusted discriminator threshold and
generates no output logic signal. Each of the last two pulses has
sufficient amplitude to produce an output logic signal. The output
signals indicated in Fig. 1 are generated when the leading edge of
the input signal crosses the discriminator threshold level.
Therefore, the time of the output response is a function of the
amplitude and rise time of the input signals. This amplitude and
rise time dependence leads to "time walk" of the output signal
relative to the beginning of the input pulse. The discriminator
output is produced earlier by pulses with larger amplitudes and
later by pulses with lower amplitudes.

Figure 2 shows three pulses that might be provided from a main
amplifier to an SCA. Only the B pulse satisfies the conditions
necessary to produce an SCA output logic signal.

Removal of the upper-level-discriminator restrictions from the
SCA allows it to be used as an integral discriminator. If the
upper-level restrictions were removed from the unit whose output is
shown in Fig. 2, both pulses B and C would be marked by logic
outputs.
Three primary modes of discriminator operation are available in
ORTEC SCAs: Integral, Normal, and Window. In the Integral mode of
operation, the SCA can function as an integral discriminator, as
indicated in the preceding paragraph. In the SCA Normal mode of
operation, the upper-level and lower-level thresholds are
independently adjustable. In the SCA Window mode, the upper-level
threshold control is used to establish a voltage level that is added
to the lower-level threshold voltage to yield the upper-level
discriminator (ULD) threshold level. Thus, when the lower-level
setting is changed, the upper-level threshold changes by the same
amount. An external voltage reference for the lower-level
discriminator (LLD) can be supplied to scan the window through a
preselected range of pulse heights.
Unlike an integral discriminator, the output logic signal from a
single-channel analyzer must be produced after the input pulse
reaches its maximum amplitude. This timing sequence must provide
sufficient time for the SCA logic circuitry to determine if the
input signal exceeded the upper-level threshold.
ORTEC provides two basic types or classifications of SCAs:
non-timing SCAs and timing SCAs. The technique used to produce the
output logic signals from an SCA determines its classification.
Non-timing units, such as the Models 550A, and 850, produce an SCA
output pulse if the input signal is within the window settings. The
output occurs when the trailing edge of the input signal recrosses
the lower-level threshold. Figure 3 shows two superimposed output
pulses from a main amplifier that meet the window requirements of
the single-channel analyzer. The output from the non-timing SCA for
each pulse is shown below the pulses. Since the linear input pulses
are referenced to the same starting time, it is clear that the
output logic signals exhibit "time walk" relative to the input
pulses.
Timing SCAs, such as the ORTEC Models 551, 552, and 590A, produce
SCA output logic signals that are precisely related in time to the
occurrence of the event being measured. This time relationship
implies that the time of occurrence of the SCA output signal is
"walk-free" or nominally independent of the amplitude of the input
signal, for a given rise time. In addition to simple counting
applications, the time-related output can be used for coincidence
measurement, pulse-shape discrimination, and other applications
where the precise time of occurrence is important.
Figure 4 shows two pulses from a main amplifier and the
response for a peak-detection single-channel analyzer such as the
Model 590A Amplifier and Timing Single-Channel Analyzer. Although
the amplitudes of the amplifier pulses differ, their peaks occur at
approximately the same time, and the SCA outputs are produced when
the peaks of the input pulses are detected.
The conventional zero-crossing technique has been widely used for
timing single-channel analyzers. This technique utilizes the
zero-crossing of the bipolar output signal from a pulse-shaping
amplifier to derive timing information, and uses the peak
amplitude of the pulse for the energy range information. Figure 5
shows two bipolar pulses provided from a main shaping amplifier.
Both pulses meet the SCA window requirements. Each output signal is
generated when the corresponding input signal crosses the baseline.
Figure 5 illustrates that the time of occurrence of the SCA output
signals is precisely related to the occurrence of the detected event
and is independent of input signal amplitude. Either
double-delay-line-shaped pulses or RC-shaped pulses may be used, but
the former provide better timing resolution. The bipolar output from
delay-line amplifiers such as the Model 460 is well suited to
zero-crossover timing with the ORTEC Model 552, because the input
signal crosses the baseline with a large slope even when the pulse
amplitude is low.
The bipolar output signal from a double-delay-line shaping
amplifier crosses the baseline at a fixed fraction that is
effectively 50% of the charge collected from the detector. Thus,
conventional zero-crossing timing can be considered as timing at a
constant fraction of the input signal amplitude. A trailing-edge
constant-fraction technique* can be used with either unipolar or
bipolar signals to derive a time-pickoff pulse after the peak time
of the signal from the shaping amplifier. This technique is
extremely useful when incorporated in timing single-channel
analyzers. Figure 6 illustrates the trailing-edge constant-fraction
technique for two unipolar input signals of identical rise times but
different amplitude. The time of occurrence of the output signals is
independent of output signal amplitudes.
The trailing-edge constant-fraction timing technique is available
with two ORTEC SCAs: Models 551 and 552.
The Model 552 can also be used as a pulse-shape analyzer. The
best known application of this technique is in the separation of the
neutron and gamma responses of some scintillators. Collection time
differences for the two types of radiation result in shape or rise
time variations in the signals from a spectroscopy amplifier. When
used with an ORTEC Time-to-Amplitude Converter, the Model 552 can
resolve these shape variations over a 200:1 dynamic range of input
signal amplitudes. The Model 552 accomplishes the shape measurement
of the input signals by evaluating the timing at two different
fractions.
The SCA function can also be applied to fast analog signals in
rising-edge constant-fraction discriminators.
*The basic circuit for implementing this technique is patented by
ORTEC, U.S. Patent No. 3,714,464.
Selection Guides |