Specifications

PERFORMANCE

DYNAMIC RANGE 200:1.

PULSE-PAIR RESOLVING TIME Output pulse width plus Delay (as selected by the front-panel Delay controls), plus 100 ns for fast NIM output or plus 200 ns for positive NIM output. Minimum resolving time for negative output 220 ns; for positive output 800 ns.

THRESHOLD TEMPERATURE INSTABILITY £±0.01%/°C of full scale, 0 to 50°C using a NIM Class A power supply (referenced to –12 V).

DISCRIMINATOR NONLINEARITY £±0.25% of full scale (integral) for both discriminators.

DELAY TEMPERATURE INSTABILITY £±0.03%/°C of full scale, 0 to 50°C.

DELAY NONLINEARITY <±2% of delay range.

WINDOW WIDTH CONSTANCY £±0.1% variation of full-scale window width over the linear range 0 to 10 V.

MINIMUM INPUT THRESHOLD 50 mV for lower-level discriminator.

TIME SHIFT vs PULSE HEIGHT (WALK) System A: Using an ORTEC Model 460 Amplifier, single delay-line mode, integrate £0.1 µs with 1-µs delay line. System B: Using an ORTEC Model 570, 571, or 572 Amplifier, unipolar output with 0.5-µs shaping time. Input from ORTEC Model 419 Pulser.

Walk (ns) Dynamic
Range
System A System B
±1.0 ±2.0 10:1
±2.5 ±4.0 50:1
±3.0 ±8.0 100:1

CONTROLS

LOWER LEVEL Front-panel 10-turn potentiometer adjustable from 0 to 10 V; when the rear-panel LL Ref mode switch is set on Int, determines the threshold setting for the lower-level discriminator. When the LL REF mode switch on the rear panel is in the EXT position, this control is ineffective.

WINDOW OR UPPER LEVEL Front-panel 10-turn potentiometer determines the window width (0 to +1 V) in the Window mode or the upper-level (0 to +10 V) threshold in the Normal mode. This control is disabled in the Integral mode.

INT/NOR/WIN Front-panel 3-position locking toggle switch selects one of three operating modes:
Integral LL sets a single-discriminator threshold (0 to +10 V) and UL is disabled.
Normal UL and LL are independently adjustable levels (0 to +10 V).
Window LL sets the baseline level (0 to + 1 0 V) and UL sets the window width (0 to +1 V).

DELAY RANGE Front-panel locking toggle switch selects delay ranges of 0.1 to 1.1 µs or 1.0 to 11 µs.

DELAY Front-panel 10-turn potentiometer for continuous adjustment of output delay over selected range. In the external strobe mode the delay control adjusts the automatic reset time from »5 µs to 50 µs.

WALK ADJUST Front-panel screwdriver adjustment for precise setting of walk compensation.

LL REF MODE Rear-panel 2-position locking toggle switch selects either the front-panel LL potentiometer or the voltage signal applied to the rear-panel LL REF EXT connector as the LL discriminator reference threshold.STROBE Rear-panel 2-position locking toggle switch selects either Internal or External source for the SCA output signal strobe function.

INPUTS

SIGNAL INPUT Front-panel dc-coupled BNC connector accepts positive unipolar or bipolar signal, 0 to +10 V linear range, ±12 V maximum; width 100 ns; 1000-W input impedance. Rear-panel ac-coupled BNC connector accepts positive unipolar or bipolar signal, 0 to +10 V linear range, ±100 V maximum; width 0.2 to 10 µs; 1000-W input impedance.

LL REF EXT When the rear-panel LL REF mode switch is on EXT, the rear-panel LL REF EXT BNC connector accepts the lower-level biasing (an input of 0 to –10 V on this connector corresponds to a range of 0 to 10 V for the lower-level discriminator setting). Input protected to ±24 V.

EXT STROBE INT When the rear-panel EXT/INT STROBE locking toggle switch is in EXT, the rear-panel EXT STROBE IN BNC connector accepts a positive NIM-standard input, nominally +5 V, 500 ns wide, to cause an output to occur from the SCA. The external strobe should be given within 5 µs (or 50 µs as determined by the front-panel Delay control) of the linear input. At the end of this period, the Model 551 resets its internal logic without producing an output signal.

OUTPUTS

SCA POS OUT Front- and rear-panel BNC connectors provide positive NIM-standard output, nominally +5 V; 500 ns wide; 10-W output impedance. For internal strobe the output occurs at the midpoint of the linear input trailing edge plus the output Delay as selected by the front-panel controls. For external strobe the output occurs at the time of strobe signal.

SCA NEG OUT Front-panel BNC connector provides fast NIM-standard output, nominally –16 mA (–800 mV on 50-W load); width £2 0 ns; rise time £5 ns; £10-W output impedance. Output occurs at the mid-point of the linear trailing edge plus the output Delay as selected by the front-panel controls.

LL OUT Rear-panel BNC connector provides positive NIM-standard output, nominally +5 V, 500 ns wide; £10-W output impedance. Output occurs as leading edge of linear input crosses the LL threshold.

UL OUT Rear-panel BNC connector provides NIM-standard output, nominally +5 V, 500 ns wide; £10-W output impedance. Output occurs as leading edge of linear input crosses the UL threshold.

ELECTRICAL AND MECHANICAL

POWER REQUIRED +12 V, 160 mA; –12 V, 110 mA; +24 V, 90 mA; –24 V, 50mA.

WEIGHT
Net
1.1 kg (2.5 lb).
Shipping
2.25 kg (5.0 lb).

DIMENSIONS NIM-standard single-width module 3.43 X 22.13 cm (1.35 X 8.714 in.) per DOE/ER-0457T.

RELATED EQUIPMENT

The Model 551 is compatible with all ORTEC amplifiers and other amplifiers having a 0 to 10 V positive, linear output range.

ORDERING INFORMATION

Model    Description
551   Timing Single-Channel Analyzer