| Home | Applications | Contact Us | ||
| Products | Service | Training | ||
|
Download PDF
|
113 Scintillation PreamplifierSpecificationsPERFORMANCE RISE TIME <60 ns. PREAMPLIFIER FALL TIME Fall time constant is designed for 50 µs, assuming a signal source impedance of 1 MW. INTEGRAL NONLINEARITY £±0.02%. TEMPERATURE COEFFICIENT ±0.01%/°C, 0 to 50°C. COUNTING RATE The gain shift of a 250-mV reference pulse is <0.25% with the application of an additional count rate of 65,000 counts/s of 200-mV random pulses. NOISE <0.1 mV rms at output. CONTROLS INPUT CAP Switch selects desired input capacity; 0, 100, 200, 500 or 1000 pF. INPUTS INPUT BNC connector; isolated for 1000 V; positive or negative polarity; input impedance is 45 pF plus the capacity selected by the front-panel switch (0, 100, 200, 500, or 1000 pF), shunted by the resistance needed to preserve a 50-µs decay time constant. TEST BNC connector accepts a pulse generator output with fast rise and slow decay to check operation of the electronics; input impedance 100 W. OUTPUT BNC connector; output impedance adjustable from 40 to 140 W. Output saturation level ±10 V into open circuit; ±5.1 V into 100-W load. Linear output ±7 V into open circuit; ±3.5 V into 100-W load. ELECTRICAL AND MECHANICAL POWER REQUIRED +24 V dc, 17 mA; -24 V dc, 17 mA. Supplied through 121-C1 cable (furnished) from any ORTEC main amplifier or preamplifier power supply. WEIGHT DIMENSIONS 4.5 x 10.2 x 15.3 cm (1.75 x 4 x 6 in.) ORDERING INFORMATION
|