ORTEC Logo

Home | Applications | Contact Us
Products | Service | Training

Download PDF

Order Online

414A Fast Coincidence

 Specifications

PERFORMANCE

PULSE PAIR RESOLUTION <100 ns on any single input; for coincidence events, <1 µs on the coincidence output.

RESOLVING TIME (2t) Continuously variable from 10 to 110 ns for coincidence signals; set by the width of the input pulse for the anticoincidence signal.

TEMPERATURE INSTABILITY 2t changes <±0.2%/°C from 0 to 50°C.

CONTROLS

RESOLVING TIME (10–110 ns) Front-panel 10-turn locking potentiometer for controlling resolving time for inputs A, B, and C over a range from 10 to 110 ns.

INPUT CONTROLS Toggle switches for using any input combination desired and for disabling input signals to the coincidence and anticoincidence circuits without input coaxial cables having to be removed.

INPUTS

COINC Front-panel BNC connectors provide 3 ac-coupled coincidence inputs (A, B, C) of positive polarity; 2-V threshold, 20-ns minimum width required; absolute maximum input 50 V; impedance >3000 W.

ANTICOINC Front-panel BNC connector provides one dc-coupled anticoincidence input (D) for inhibiting coincidence output; +2 V threshold, 20-ns minimum width required; absolute maximum input 50 V; impedance >3000 W.

OUTPUTS

OUTPUT Two separate buffered coincidence output signals through front-panel BNC connectors provide positive pulses ³500 ns wide with 5-V minimum amplitude; ac-coupled with <10-W impedance; monitored through oscilloscope test points on front panel.

ELECTRICAL AND MECHANICAL

POWER REQUIRED The Model 414A derives its power from a standard NIM bin/power supply. The power required is +24 V, 30 mA; –24 V, 30 mA; +12 V, 120 mA; and –12 V, 85 mA.

WEIGHT
Net
1.09 kg (2.4 lb).
Shipping
2.0 kg (4.4 lb).

DIMENSIONS NIM-standard double-width module 6.90 X 22.13 cm (2.70 X 8.714 in.) per DOE/ER-0457T.

ORDERING INFORMATION

Model    Description
414A   Fast Coincidence