|
Case 1: ∆t << Td
Cascaded Dead Times:
As explained in the introduction to the Fast Timing Discriminators
section, the dead time experienced in the counting chain is
typically composed of two cascaded components, Te and Tne.
Te is the extending dead time caused by the duration of
the analog signal from the detector at the noise threshold of the
timing (or counting) discriminator. It is an extending dead time
because a second analog pulse occuring during a preceeding pulse
extends the dead time by one pulse width from the arrival time of
the second pulse. The non-extending dead time, Tne, can
be caused by the pulse width of the discriminator output driver, or
it can be a longer dead time contributed by a circuit in the MCS or
time digitizer. Sufficient accuracy1,2 will be achieved
if one chooses the longer of these two dead times to represent Tne.
A second pulse occurring during Tne is ignored and does not affect
the dead time. It is convenient to define the approximate dead time
in the system as
Td ≈ Te
+ U(TneTe) (TneTe)
(1)
where U(TneTe)
is a unit step function defined by
U(TneTe)
= 1 for Tne > Te
= 0 for Tne ≤ Te
(2)
Under that definition, the equations
for Case 1 are valid if the quantization interval,
Δt, is insignificant compared to Td.
This is the practical situation encountered in the Model 9308
picosecond TIME ANALYZER. For the Model 9308, Td
≈ 45 ns and the maximum size of the bin
width, Δt, is 2.5 ns.
Presume a time digitizer that has summed the repetitive spectra from
n start triggers. The counts in the ith bin of the resulting
spectrum (after suffering dead time losses) are defined to be qi,
and the time, t, is related to the bin number by
t = i
Δt
(3)
By analogy to equation (3) it is
convenient to define the quantized dead times, te,
tne, and td,
by equations (4).
Te =
te Δt
(4a)
Tne = tne
Δt
(4b)
Td = td
Δt
(4c)
Note that i, te,
tne and td
are all rounded to integer values.
The number of counts that would have been recorded in bin i if the
dead time were zero is defined to be Qi. The distorted
spectrum recorded in the measurement is represented by qi,
whereas Qi is the undistorted spectrum that is sought.
When the counting rates are low enough to yield single-ion or
single-photon counting, one can apply statistical sampling theory.
Poisson Statistics can also be applied directly, provided the dead
time losses are negligible3.
In equation (5), qi /n is the probability of recording an
event in the ith bin during a single pass through the time span. It
is composed of three probabilities4, as described in the
right hand side of the equation.
Cascaded Dead Time Equation:
(5)
The first term, Qi /n is the
probability that an event will arrive at the detector at a time
suitable to be categorized in bin i. In order to be recognized as
distinct from previous analog pulses there must be no pulses
arriving at the detector in the time interval te
preceeding the pulse for bin i. That probability is given by the
exponential term in equation (5). If a pulse had been counted by
triggering the non-extending dead time in the time interval
tne preceeding the pulse for
bin i, the pulse for bin i would be lost. Consequently, the term in
the square brackets is the probability that no pulses are recorded
in the preceeding time interval tne.
Note that the sum stops at j = i te
1 because the exponential term already guarantees no pulses occured
in the time interval from j = i te
to i 1.
Equation (5) can be rearranged to get the formula for computing the
corrected spectrum, Qi, from the distorted spectrum, qi.
(6)
One applies the correction algorithm by
starting at bin i = 0, while presuming that qi, qj,
and Qj are all zero for negative values of i and j. For
each i, the value Qi is calculated from equation (6)
using the recorded values qi and qj along with
the Qj values calculated for the previous values of i.
This correction calculation is applied bin by bin until the maximum
bin in the spectrum has been treated. At that point the list of Qi
values is the corrected spectrum.
If there truly were no counts to be detected for negative values of
i, then the Qi data near i = 0 will represent the true
spectrum before dead time losses. If the detector was actually
responding to events for negative values of i, then Qi
will be underestimated until the bin number exceeds several times
td. Frequently, this
shortcoming can be eliminated by inserting a coaxial cable delay of
the appropriate length between the detector and the stop input on
the time digitizer.
Single, Extending Dead Time:
For a system where the detector pulses are longer than any other
dead times in the system (te >
tne), equation (5) simplifies
to the equation for a single, extending dead time4.
i 1
qi = Qi exp {Σ Qj
/n } (7)
j = i te
Single, Non-extending Dead Time:
The other extreme is a system in which detector pulse widths are
negligible compared to the non-extending dead time in the MCS or
time digitizer. In that case, equation (6) simplifies to4,5
(8)
Accuracy of the Dead Time
Correction:
It can be demonstrated by substituting known values into equations
(6), (7) and (8) that all three equations yield predictions of Qi
/ qi that are within 1% of each other provided Qi
/ qi < 1.15 (i.e., a dead time correction < 15%), and
provided td is substituted for the single dead times in equations
(7) and (8). This allows a simpler correction algorithm to be
implemented using equation (8). In fact, the algorithm using
equation (8) can start at the maximum value of i and proceed towards
i = 0, while replacing qi with Qi in the data
file. This is the procedure used in the Model 9308 software.
Without the dead time correction algorithm, one would have to limit
the counting rate to achieve <1% dead time losses in order to limit
the spectrum distortion to <1%. By applying the dead time correction
algorithm, one can typically operate at a factor of 10 higher dead
time loss, while still achieving <1% spectrum distortion. This
implies a factor of 10 higher data rates. However, the accuracy of
the correction is limited by the factors discussed next.
If the time spectrum is constant across all bins, it is easy to show
that a 10% error in the assumed value for the dead time in equations
(6), (7), or (8) will lead to < 1% error in the corrected counts if
Qi / qi < 1.10. A more serious case is a
narrow peak centered at bin i, and preceeded by an intense, narrow
peak centered at bin j = i td.
A small error in the presumed value for td
can result in either including or excluding the peak at bin j in the
dead time corrections. This can make a large difference in the dead
time correction applied to the peak at bin i. An additional issue is
the error in rounding off the presumed dead time to the nearest
integer value. This leads to round-off errors at the two limits of
the sums in the equations. That effect can be restricted to an error
<1% if one ensures that qj /n < 0.005 for all j. Clearly,
it is important to use an accurately measured dead time in the
correction formula.
By applying basic probability theory, it can be shown that the
statistical variance in the recorded counts qi is given
by4
σqi2 = qi
(1 qi /n)
(9)
≈ qi
for qi /n << 1
Moreover, for the sum of the recorded
counts in any number of bins, such as
i 1
M = Σ
qj
(10)
j = i t
the statistical variance in the sum is
i 1
σM2 = M =
Σ qj
(11)
j = i t
A straight forward
propagation-of-errors computation predicts the statistical variance
in the corrected counts Qi calculated from equation (8)
to be4,5
σqi2 = Qi
ki
(12)
where ki defines the
magnification factor arising from the variances in the qi
and qj in equation (8), i.e.,
ki = (Qi
/ qi) + (Qi / qi)2 [(Qi
/ qi) 1] (qi /n)
(13)
The effect of ki is small
for dead time corrections Qi / qi < 1.15, but
the second term of ki escalates rapidly for higher dead
time corrections.5
1Jφrg W. Mόller, Nucl. Instr. Meth. 112, (1973), 4757,
Fig. 3.
2D.R. Beaman, et al., J. of Physics E: Sci. Instr.
(1972), 5, 767776.
3Ron Jenkins, R.W. Gould, and Dale Gedcke, Quantitative
X-Ray Spectrometry, Marcel Dekker, New York, (first edition), 1981,
Chap. 4.
4D.A. Gedcke, Development notes and private
communication, Nov.Dec. 1996.
5P.B. Coates, Rev. Sci. Instrum. 63 (3), March 1992,
2084. |