ORTEC Logo

 
Home | Applications | Contact Us
Products | Service | Training

Specifications

Download PDF

Order Online

863-fp.jpg (16426 bytes)

9302 AMPLIFIER DISCRIMINATOR

  • Fast amplifier and discriminator for photon, electron, or ion counting applications
  • 100-MHz counting rate capability
  • Includes rate monitor facility

The ORTEC Model 9302 is a fast Amplifier and Discriminator in a single- wide NIM-standard module designed for use with photomultiplier or electron multiplier tubes in photon, ion, or electron counting applications. Features of the Model 9302 include a wideband, high-gain amplifier and an integral discriminator capable of counting rates up to 100 MHz.

Incorporated into the module is a rate monitor system. This facility can be used to perform external switching when the pulse counting rate is modulated. At counting rates equal to or above those selected by the monitor rate discriminator, a voltage of >+2 will be present at the signal monitor connector. At counting rates below the rate selected, 0 V will be present at the connector. The threshold value is selected by the front-panel switch.

The amplifier section of the Model 9302 is ac-coupled with input and output impedances of 50 W. The amplifier has a rise time of typically 3 ns at a gain setting of 200 and <2 ns when the gain is set at 20.

The discriminator section of the Model 9302 is a leading-edge discriminator whose level can be adjusted through a range from 50 mV to 1 V.

ORDERING INFORMATION

Model   Description
9302   Amplifier Discriminator