  Selection Guide for ADCs

These charts are intended as selection guides only. For complete and precise specifications, consult the data sheet for each model.

Model AD114 CAMAC 16k ADC AD413A CAMAC Quad 8K ADC
Primary Applications Coincidence or singles measurements with multiple detectors. Fast readout via FERAbus. Live-Time clock. Coincidence or singles measurements with multiple detectors. Fast readout via FERAbus.
Package Width CAMAC-2 CAMAC-2
Number of Channels 16128 8064
Input Range 0 to +10 V 0 to +10 V
Number of Inputs 1 4
Gating Inputs Master gate, local gate, PUR Master gate, individual gates, individual PUR
Type Peak amplitude sensing; successive-approximation ADC with sliding scale linearization Peak amplitude sensing; successive-approximation ADC with sliding scale linearization
Conversion Time 5 µs 6 µs per active input
Digital Offset No No
Live-Time Clock Yes No
Readout CAMAC or fast FERAbus CAMAC or fast FERAbus
Compatible Memories ORTEC HM413 (Histogram)
LeCroy 4302 (FIFO)
CES HSM8170 (FIFO)
ORTEC HM413 (Histogram)
LeCroy 4302 (FIFO)
CES HSM8170 (FIFO)
Multi-Parameter Software Kmax from Sparrow Kmax from Sparrow