ORTEC Logo

 
Home | Applications | Contact Us
Products | Service | Training

Dimensional Drawings
 

X-COOLER vs. LN2 Specs

 
Download
Brochure
(KB PDF)
 
Visit MMR.com  makers
of the
X-COOLER II technology
X-COOLERII

Specifications

RESOLUTION

Detector specifications are warranted to no degradation above 500 keV and less than 10% degradation of the warranted and published LN2 specifications at less than 500 keV. Planar detector specifications are warranted to no degradation above 500 keV and less than 20% degradation of the warranted and published LN2 specifications at less than 500 keV.

MECHANICAL

Dimensions

Compressor: 12.5" W x 12.5" D x 11" H (31.8 cm x 31.8 cm x 28 cm)

Cold Head Length with Detector Capsule
Standard: 23.25" (59 cm)|
Oversize: 24.5" (62.3 cm)
Ultra: 25.5" (64.8 cm)

Weight

Compressor: 36 lbs. (16.4 kg)

Cold Head:  11 lbs. (5 kg) not including capsule

Noise: Less than 70 dB at 1 meter

Input Power: 110–120 V ac, 57–63 Hz or 220–240 V ac, 47–53 Hz (input power must be selected by choosing the appropriate model number, see ordering information)

Power Consumption:  <500 W during initial cool down. <400 W during normal operation.

Connector Hose:  10 ft. gas hose connected between compressor box and cold head assembly is included. Longer lengths available on special request. Contact the factory for more information and pricing on other hoses.

Refrigerant:  Mixed gas, CFC-free refrigerant. MSDS available on request.

CE Approved

ENVIRONMENTAL

Ambient Temperature:  5-30°C (40-85°F)

Relative Humidity 5-95% non-condensing

ORDERING INFORMATION
 

Model

Description

CFG-X-COOL-II-115 X-COOLER II with PopTop connector using 110–120 V ac, 60 Hz Input Power
CFG-X-COOL-II-230 X-COOLER II with PopTop connector using 220–240 V ac, 50 Hz Input Power
X-COOL-STAND Omnidirectional Cold Head Stand (ideal when replacing 30 L dewar dipstick type cryostats)
X-COOL-RACKMOUNT Cold Head Rackmount Bracket
X-COOL-PK-110
(or -220)
X-COOLER Pump-out Kit
CRYOSECURE Programmable Compressor Power Controller (ensures that detector is fully warmed before commencing cooldown after power failure)
X-COOL-UPS-115 30 minute backup power system (1440 VA) for maintaining Mains power to the X-COOLER II during brief power outages. Input/Output power: 110–120 V ac, 60 Hz.
X-COOL-UPS-230 30 minute backup power system (1440 VA) for maintaining Mains power to the X-COOLER II during brief power outages. Input/Output power: 220–240 V ac, 50 Hz.
X-COOL-FILTER Replacement Filters (pkg. of 12).