SMART-1 and Detector Interface Modules (DIMs)
SMART-1 is a significant advance in the operation of HPGe detectors
and is a major enhancement to data and chain-of-custody integrity.
It is supported by the digiDART, DSPEC jr 2.0, and DSPEC pro
revolutionary, small sized, low power, DIGITAL nuclear multichannel
analyzer systems. Both instruments are designed primarily with the
needs of HPGe-detector gamma-ray spectrometry in mind. The SMART-1
HPGe detector has the high voltage included, so none of the
instruments require an external high-voltage power supply. The
SMART-1 HPGe detector monitors and reports on vital system
functions. Also, the SMART-1 identification serial number can be
read by the instrument. For more security it can save authentication
codes and report the code at a later time.
ORTEC SMART-1 HPGe Detectors monitor the following conditions:
Preamplifier +24 and +12 V values (read-only)
Detector element temperature (read-only)
Detector high voltage value (read-only)
Detector high voltage state (on/off)
Detector overload state (read-only)
Detector HV shutdown state (read-only)
Detector serial number (read-only)
Detector authentication code (read/write)
These parameters are displayed as required on the LCD display of
the attached instrument or computer, local or remote.
SMART-1 Physical
The SMART-1 is housed in a rugged ABS molded plastic enclosure. The
SMART-1 is permanently attached to the detector endcap via a
molded-strain-relieved sealed cable. This eliminates the possibility
that the detector will suffer severe damage from moisture leaking
into high-voltage connectors. The SMART-1 can be positioned in any
convenient place and does not interfere with shielding or other
mounting hardware.
The Detector Interface Module (DIM)
For older, non-ORTEC or non-SMART-1 detectors, the DIM is used to
supply the HV for the detector. The DIM is provided in the same
convenient housing as the SMART-1, but with flying leads (HV, Bias
Supply Shutdown, Energy Output, Preamplifier Power, and Inhibit
Output) to connect to the HPGe detector. The DIM does NOT provide
SMART-1 functionality to a non-SMART-1 detector. It supplies HV and
preamplifier power as well as providing warm-up protection for he
HPGe detector.

The following DIMs support non-SMART-1 detector systems:
- DIM-POSGE For any non-SMART-1 positive-bias, HPGe
detector. Includes short leads with mating connectors for
connecting to existing detector cabling.
- DIM-NEGGE For any non-SMART-1 negative-bias, HPGe
detector. Includes short leads with mating connectors for
connecting to existing detector cabling.
- DIM-POSNAI For any positive-bias NaI detector.
Includes short leads with mating connectors for connecting to
existing detector cabling.
Specifications
Dimensions: The DIM is shaped for easy mounting; the
maximum dimensions (including mounting lugs) are 12 cm x 8.2 cm x
3.3 cm (4.7 in. x 3.2 in. x 1.3 in.)
Weight: <240 gm (0.5 lb.)
Performance
- DIM-POSGE Output 0 +5 kV, 100 MΩ
impedance
- DIM-NEGGE Output 0 -5 kV, 100 MΩ
impedance
- DIM-POSNAI 0 +1800 V, 600 ľA below 1000 V, 300 ľA maximum
above 1000 V
DIM Connectors
Multi-Pin connector for attachment to instrument (mating cable is
provided with instrument)
DIM-POSGE and DIM-NEGGE (these connectors will mate directly with
most standard HPGe detectors)
- Preamplifier Power: 9-PIN D female 1 W maximum (+12 V, 12 V,
+24 V, 24 V, 2 GND)
- Analog In: BNC male
- Bias Shutdown In: BNC male
- Bias Out: SHV (HV) female
- TRP Inhibit Out/Coincidence Gate In: BNC male (function is
software selected)
DIM-POSNAI
- Preamplifier Power: 9-PIN D female 1 W maximum (+12 V ,12 V,
+24 V, 24 V, 2 GND)
- Analog In: BNC male
- Bias Out: SHV (HV) female
- Coincidence Gate In: BNC male
DIM-296 Option for 14 Pin PMT NaI Detectors
The DIM-296 option is an alternative to the DIM-POSNAI for NaI
detectors with a standard 14-Pin base. The DIM-296 is a 296
ScintiPack with a single cable connection, ten feet in length for
direct connection to the instrument.
- For scintillation detectors employing 10-stage PMTs that fit
standard 14-pin sockets
- Convenient, single-cable connection direct to instrument
- Internal, adjustable, high-voltage bias supply eliminates
high-voltage cables
- Active bias network minimizes peak shifts at high counting
rates
- Low power consumption (240 mW) for portable applications

|