Germanium Detector Stocklist

GLP Series Planar Germanium (HPGe) Low-Energy Photon Detectors (in PopTop Capsule)

  • Excellent energy resolution in the 3 to 300 keV range
  • Active area from 30 to 1000 mm2
  • PopTop flexibility
  • Unequaled timing performance
  • Detectors larger than 36-mm diameter available
  • Available with special feedback resistor for high-rate applications
  • POF preamplifier option for superior energy resolution and high count rate at low energies
b23.jpg (19590 bytes)

Figure 23. Energy Resolution at High Count Rate
for Detector D-256.

b24.jpg (20660 bytes)

Figure 24. 16-mm Diameter, 10-mm Deep GLP Detector.

The ORTEC GLP Series Planar Low-Energy Photon Spectrometer (LEPS) is a small-area, high-purity germanium photon spectrometer for use in applications over the energy range from 3 to ~300 keV.

Available in diameters from 6 to 36 mm, the LEPS offers exceptional energy resolution for low and intermediate energies. At low energies in nuclear structure physics GLP detectors are irreplaceable because of their excellent timing performance (see Table 7). At energy rates exceeding the values shown in Table 2 (for example, in Safeguards applications), good performance can be obtained by using a lower value feedback resistor (Fig. 23).

A cross sectional drawing of a 16-mm LEPS is shown in Figure 24.

 

 

 

Ordering Information

GLP Series Planar, Germanium (HPGe) Low-Energy Photon Spectrometer in PopTop Capsule
A complete detector assembly requires a cryostat and a dewar.
 

Model No.

Active
Diameter
(mm)

Dimensions

Energy Resolution FWHM**

Area (mm2)

Depth* (mm)

at 5.9 keV

at 122 keV

GLP-06165/05P4

6

28

5

165

480

GLP-10180/07P4

10

80

7

180

485

GLP-16195/10P4

16

200

10

205

495

GLP-25325/10P4

25

500

10

325

550

GLP-25300/13P4

25

500

13

300

545

GLP-32355/10P4

32

800

10

355

580

GLP-32340/13P4

32

800

13

340

570

GLP-36385/10P4

36

1000

10

385

595

GLP-36360/13P4

36

1000

13

360

585


* Depletion depths are minimums. If a specific maximum depth is required, please specify.

**Resolution performance specifications shown are applicable to standard resistor feedback systems.

† Supplied with 0.005-inch-thick Be window.

‡ Supplied with 0.010-inch-thick Be window.

For high rate applications, contact the factory for results obtainable with a POF preamplifier.