IGLET Germanium (HPGe) Detector for Low-Energy X Rays

  • Unmatched energy resolution at low and high count rates
  • Special detector geometry allows dense packing
  • Ideal for most XAFS applications
  • X-ray spectroscopy down to 3 keV
  • Available in custom XAFS arrays and a variety of streamline cryostats, including the JIMseries — ideal for spectroscopy on small targets
  • Ideal for higher energy PIXE applications
The IGLET detector geometry results in substantially better resolution than is achievable with Germanium planar detectors. By virtue of a proprietary detector-element geometry and an advanced reset-feedback preamplifier, the IGLET offers unmatched performance from 3 keV to 30 keV, the energy range for XAFS measurements at synchrotron light sources. Furthermore, the IGLET has unmatched warranted performance at 100,000 cps, thus ensuring full exploitation of the extreme brightness of x-ray beams at the newest facilities. Some custom-design XAFS arrays.

Entrance Window

A thin Be window is provided which is 80 to 90% transparent to 3-keV x rays. The factory may be contacted to discuss alternate choices of appropriate entrance window for your particular application.

Individual Detector or Multi-Channel Arrays

The IGLET is available either as an individual detector in an individual cryostat or in multi-channel arrays.

Ordering Information

Individual IGLET detectors can be supplied in ORTEC "streamline," but not PopTop, cryostats. The choices are vertical dipstick (SV), horizontal dipstick (SH), side-looking "bucket" (SL), down-looking "bucket" (SD), and Gamma Gage (GG). For spectroscopy on extremely small targets JIM series cryostats with their small diameter are the best choice. There are several choices of liquid-nitrogen dewar size. Dimensional information for the cryostats and dewars.

IGLET Germanium (HPGe) Detector Element for Low-Energy X Rays
To order a complete IGLET Detector, you must order a detector element, a cryostat, and a dewar (or cryostat/dewar assembly).

Model Active
Diameter
(mm)
Active
Depth
(mm)
Warranted Resolution Be Window
Thickness
(µm)
@5.9 keV, 10 µs
1000 cps
@5.9 keV, 0.5 µs
100,000 cps
IGLET-06135 6 6 135 eV 250 eV 13
IGLET-11145 11 10 145 eV 270 eV 25
IGLET-16160 16 10 160 eV 280 eV 50