Germanium Detector Stocklist

 

High-Count-Rate Detectors

High Count Rates with Germanium (HPGe) Detectors in the Energy Range Above 300 keV

The subject of high count rate spectroscopy with germanium detectors and analog electronics is thoroughly discussed in the paper  High-Count-Rate Spectroscopy with Ge Detectors: Quantitative Evaluation of the Performance of High-Rate Systems.

High Count Rate in the Energy Range 30 keV to 300 keV

GLP planar detectors, 16, 25, 32, or 36 mm in diameter with appropriately selected feedback resistors, are the instruments of choice for studying the complex gamma spectra generated by highly enriched materials in the 80- to 200-keV region.

The value of the feedback resistor (RF) for these detectors is selected in such a way that the energy rate is maximized without a substantial degradation of the energy resolution.

a31.jpg (18623 bytes)
Figure 31. Performance of a high-counr-rate 500-mm2
active area planar HPGe detector at 122 keV.

These detectors are typically characterized by their high-count-rate performance at 122 keV. Figure 31 shows the performance of 500-mm2 active area planar detectors built for these applications.

High Count Rate in the Energy Range <30 keV

In the energy range up to 60 keV the detectors of choice are the IGLET and the IGLET-X. For such detectors, both at low and at high count rates, the preamplifier of choice is the pulsed optical feedback, POF preamplifier .