Detector Microphonics

After more than 30 years of germanium detector production the phenomenon of microphonics is still not well understood. A back-of-the-envelope calculation leads to the false conclusion that no germanium detector will ever operate. For example, consider a metal part, such as the cup that holds the detector, which has a small, but non-zero capacitance with respect to the FET gate. Assume that sound waves, such as from a voice, induce a variation of merely 0.5 femtofarads in the value of this capacitance; the result would be a signal equivalent to 10 keV!

Although there is no IEEE standard on the measurement of the extent of microphonics, considerable work has been done in this field:

1) Special design: ORTEC has always been at the forefront in this field, for example production of a rugged detector designed for the U.S. Navy (Ref. 11). ORTEC has also provided arrays of germanium detectors for helicopter aerial surveillance.

2) Proper electronic setup: As the microphonics spectrum is primarily in the few kcps range, a high pass filter (shorter amplifier time constants and baseline restorer "on") will often improve detector performance.

3) Vibration decoupling: Users typically obtain improvement by using soft foam rubber around and under the detector.