ULTRA CAMIon-Implanted-Silicon Detectors for Continuous Air Monitoring
Continuous air monitoring (CAM) instrumentation is used for counting, or rough-spectroscopy, on alpha particles from filters used with continuous air monitors. ULTRA CAM detectors set a new standard for silicon detectors for this purpose. Since CAM instruments must work in air, exterior light, and under uncontrolled ambient conditions, the entrance contact of the ULTRA CAM detectors is coated with both an aluminum evaporation and a thin polymer film for protection against adverse environmental conditions, such as high humidity.ULTRA CAM detectors have active areas ranging from 300 to 2000 mm2. They operate at low voltage (+15 to 24 volts). This means that a separate HV supply is not required; the NIM +24 volt supply can be used for application of bias. The alpha resolution that can be achieved is determined solely by the distance the alpha particles travel in air before reaching the detector. A U-CAM-450 in vacuum yielded an energy resolution on 5.47 MeV alphas of 35 keV FWHM.The detectors are normally supplied in "B" mount, with a rear Microdot connector. Other mounting is available on request. (See mount descriptions.) Ordering Information Supplied with B Mount unless otherwise specified.
*Deeper detectors available on special order |