R Series Ruggedized Partially Depleted Silicon Detectors
Main Application: Charged-particle spectroscopy in vacuum or in air
with ambient light; cleanable.
Supplied with B Mount unless otherwise specified. (See mount descriptions.)
|
Active
Area
(mm2) |
Guaranteed
Maximum
Resolution
(keV) |
Minimum
Depletion
Depth
100 µm
Model No. |
Minimum
Depletion Depth
300 µm
Model No. |
Guaranteed
Maximum
Resolution
(keV)** |
Minimum
Depletion Depth
500 µm
Model No. |
|
a |
b |
a |
b |
| 50 |
15 |
7 |
R-015-050-100 |
R-015-050-300 |
15 |
7 |
R-015-050-500 |
| |
17 |
8 |
R-017-050-100 |
R-017-050-300 |
17 |
8 |
R-017-050-500 |
| 150 |
16 |
10 |
R-016-150-100 |
R-016-150-300 |
16 |
10 |
R-016-150-500 |
| |
19 |
11 |
R-019-150-100 |
R-019-150-300 |
18 |
11 |
R-018-150-500 |
| 300 |
19 |
13 |
R-019-300-100 |
R-019-300-300 |
19 |
13 |
R-019-300-500 |
| |
21 |
15 |
R-021-300-100 |
R-021-300-300 |
21 |
15 |
R-021-300-500 |
| 450 |
20 |
14 |
R-020-450-100 |
R-020-450-300 |
20 |
14 |
R-020-450-500 |
| |
24 |
19 |
R-024-450-100 |
R-024-450-300 |
24 |
18 |
R-024-450-500 |
| 600 |
25 |
17 |
R-025-600-100 |
R-025-600-300 |
25 |
19 |
R-025-600-500 |
| |
33 |
28 |
R-033-600-100 |
R-033-600-300 |
33 |
29 |
R-033-600-500 |
| 900 |
30 |
25 |
R-030-900-100 |
R-030-900-300 |
30 |
25 |
R-030-900-500 |
| |
40 |
35 |
R-040-900-100 |
R-040-600-300 |
53 |
48 |
R-053-900-500 |
| 2000 |
50 |
45 |
R-050-2000-100 |
R-050-2000-300 |
50 |
45 |
R-050-2000-500 |
| |
60 |
55 |
R-060-2000-100 |
R-060-2000-300 |
80 |
75 |
R-080-2000-500 |
Supplied in B Mount unless specified otherwise by the appropriate letter prefix.
Other areas, depths, and configurations available on special order.
First three digits of Model No. indicate total system resolution FWHM for 241AM,
5.486-MeV alphas, using standard ORTEC electronics and 0.5-µs shaping time constants.
Beta resolution approximated by pulse generator width FWHM.
|