Detective
and Detective-100
HPGe-Based Portable Nuclide Identifier
Specifications
OPERATION MODES
SEARCH
Scanning mode for location of radioactive sources, with audio alert
using an external ear piece. Both neutron and gamma search is simultaneous;
5 speed settings.
IDENTIFY-Gamma
Proprietary scheme for identification and classification of radionuclides
as:
Industrial: Including
57Co, 60Co, 133Ba, 137Cs,
192Ir, 241Am, 75Se
Medical: Including 18F, 67Ga, 99mTc,
111In, 123I, 131I, 133Xe,
201Tl
Natural (NORM): Including 40K, 226Ra, 232Th,
238U
Nuclear: Including 233U, 235U, 237Np,
239Pu, 252Cf
These classifications are based on an internal,
fixed library according to ANSI N42.34. Customized libraries for specific
applications can be supplied by special order.
Dose Rate
Visual overrange indication and continuous audible alarm at dose
rates >10,000 µSv/hr.
DETECTORS
Internal HPGe Detector
Detective Crystal Nominal Dimensions: 50 mm diameter x 30 mm deep.
Detective-100 Crystal Nominal Dimensions: 65 mm diameter x 50 mm deep.
P-type high-purity germanium. Coaxial construction.
Cooler: Hymatic SAX101-002 high reliability, low power Stirling Cooler.
Cooler design life >5 years continuous running. Dual piston design,
1 W nominal lift at 100°K.
Digital Noise Suppression: "LFR Filter" ORTEC Patent Pending.
Gamma Dose Rate Detector
Two detectors determine the gamma dose rate over a wide range from
<0.05 µSv/h to >10000 µSv/h, a dose-rate range of around six decades.
For low dose rates, below ~20 µSv/h, the dose rate is determined from
the Ge detector spectrum. For dose rates above this value, the internal
compensated GM tube is used. Detective switches between the two automatically:
Dose rate uncertainty ≤50% to +100%; continuous audible alarm at dose
rates >10,000 µSv/h.
DIGITAL MCA AND DATA PROCESSOR
Display: 240 x 160 pixel backlit LCD.
Keypad: 21 numeric and function keys. Dedicated
large single key for initiation of ID, Search and MCA display modes.
-
Simple to use menu operation
-
Digital MCA with internal storage of multiple
spectral data
-
Maximum number of stored spectra >40.
-
8k Channel Conversion gain.
-
Monitoring of vital system functions:
Instrument Battery life remaining
System DC voltages
Detector Bias for both HPGe and GM
HPGe Crystal Temperature
Instrument is supplied factory pre-calibrated and
adjusted. A recalibration function allows correct performance to be
verified and adjusted using a small 137Cs source.
DETECTIVE DISPLAYS AND MENUS
Main Screen
Gamma Count Rate Bar Graph 20 kcps full
scale. Should the count rate exceed 20 kcps, bar is completely black
and the rate line flashes.
Dose Rate Bar Graph 10 mSv/hr full scale,
flashes on overrange.
Status Lines:
WARNING!! High Dose Rate — Displayed when Dose rate exceeds 10 mSv/hr
Detector is Warm
— Displayed when crystal temperature is above working limit
Bias Supply Error
— Displayed if any power supply is bad
WARNING!! Low
Battery
Pressing Search starts a search mode acquisition
Pressing ID starts an identify operation
Pressing Backlight turns the light on or
off
Pressing Display brings up the spectral
display
Pressing Menu brings up the Main Menu
Search
Search starts when the "Search" button is pressed on the keypad.
In search mode, a "strip chart" is drawn across the screen showing count
rate vs. time on a log scale. 5 speed settings are available. If dose
rate is beyond the maximum, the bar is black and flashes. The data may
be saved.
Identify
When ID is pressed, data is cleared and acquisition starts.
Nuclides are divided into 5 basic types (Nuclear,
Industrial, Medical, Natural, and Other).
Pressing Menu goes to the following options:
1 - View Found Nuclides returns confirmed identified nuclides
2 - View Suspected Nuclides
3 - View Top Lines returns the "Top 10 List"
4 - Save Saves the spectrum
Display brings up the spectral display. The spectrum
may be manipulated via the arrow keys and various accelerator keys for
cursor movement. Energy and channel contents are displayed with the
spectrum.
Main Menu
1 - Administration Menu
2 - Calibrate
3 - Audio Settings
4 - View Diagnostic Status
1 - Administration Menu
Password Protected. Once verified, the spectrum screen is shown.
Spectrum may be manipulated with Arrow keys and via zoom and jump "hot
keys".
The Administration Menu has the following items:
1 - View Status
2 - General Settings
3 - Change Clicker Sensitivity
4 - Store & Retrieve Spectra
5 - Cooler Settings
6 - Search Mode Settings
7 - Change Dose Units
8 - Auto-Calibration Setup
9 - Exit Administrator Mode
2 - Calibrate
Calibrate looks for the 137Cs peak in a range of 650
keV–680 keV. Once the peak is located, and there are at least 1000 counts
in the peak (net area), a new fine gain is calculated. This is executed
automatically when on the docking station.
3 - Audio Settings
Allows setting various parameters associated with the speaker and
headphone.
4 - View Diagnostic Status
Reports instrument status, for example:
+12V
OK
–12V
OK
+3.3V
OK
Detector Temp
Body Temp OK
Cold Tip Temp OK
Cooler Drive OK
HV Volt
OK
PHYSICAL SPECIFICATIONS
Detective
Maximum Overall Dimensions (including handle,
Ge detector endcap and shock absorbers): 37.3 cm L x 16 cm W x 32 cm
H (14.7" L x 6.3" W x 12.6" H)
Weight: 22.9 lb (10.39 kg)
Detective-100
Maximum Overall Dimensions (including handle,
Ge detector endcap and shock absorbers): 39.4 cm L
x 16.3 cm W x 32 cm H (15.5" L x 6.55" W x 12.6" H)
Weight: 23.3 lb (10.65 kg)
Internal Battery Life: >3 hours at
25°C with a cold detector on fully charged internal battery; battery
lifetime may be extended indefinitely by the use of external battery
packs which are available in "battery belt" formats. The unit is expected
to be kept running once cold.
Input Power: 10–17 V dc 30 Watt or via auto-sensing
Mains powered Battery Charger.
Temperature Operation Range/Humidity:
–15°C to +50°C, Relative Humidity <90% at 35°C, non-condensing.
Communications Ports
USB 1.1 connection provided at the rear panel, protected by dust
cover.
External Power and DC in.
Cool Down Time: The high reliability cooler
is designed for continuous operation. Initial cool down and holding
ready for use is normally achieved by attachment to the supplied docking
station, which may be powered from DC or Mains, and both holds and charges
the internal battery. The cooler life is expected to exceed 50,000 hours
continuous operation. Initial cool down time depends on ambient temperature,
but is typically 12 hours.
ORDERING INFORMATION
|
Model
|
Description
|
|
DETECTIVE
|
Includes Detective HPGe Identifier, docking station with calibration
check source, universal mains supply, automobile power cables
and shoulder strap.
|
|
DET-PKG-1
|
Includes Detective HPGe Identifier, docking station with calibration
check source, universal mains supply, automobile power cables,
shoulder strap, A65-B32 software and large wheeled transport
case.
|
|
DETECTIVE-PAC
|
Includes Detective HPGe Identifier, ac/dc power adapter/charger
(PAC), automobile power cables, and shoulder strap. (No Source.)
|
|
DET-PAC-PKG-1
|
Includes Detective HPGe Identifier, ac/dc power adapter/charger
(PAC), automobile power cables, shoulder strap, A65-B32 software
and compact wheeled transport case.
|
|
DETECTIVE-100
|
Includes Detective-100 HPGe Identifier, docking station with
calibration check source, universal mains supply, automobile
power cables and shoulder strap.
|
|
DET-100-PKG-1
|
Includes Detective-100 HPGe Identifier, docking station with
calibration check source, universal mains supply, automobile
power cables, shoulder strap, A65-B32 software and large wheeled
transport case.
|
|
DET-100-PAC
|
Includes Detective-100 HPGe Identifier, ac/dc power adapter/charger
(PAC), automobile power cables, and shoulder strap. (No Source.)
|
|
DET-100-PAC-PKG-1
|
Includes Detective-100 HPGe Identifier, ac/dc power adapter/charger
(PAC), automobile power cables, shoulder strap, A65-B32 software
and compact wheeled transport case.
|
Note on Calibration Check Source
The calibration check source is internal to the docking station.
It comprises a solid, sealed, ~0.25-µCi (9250-Bq) 137Cs calibration
source housed in the source holder on the front of the docking station.
This source is exempt under U.S. and European regulations and is only
supplied pre-installed by ORTEC (not separately). It is not recommended
that it be removed, and ORTEC will not accept any responsibility for
its removal. If this source is NOT required with the instrument because
of local issues, this must be stated at time of order. THE SOURCE CANNOT
BE SUPPLIED SEPARATE FROM THE DOCKING STATION.
Options
A separate data sheet covering Detective Options and accessories
is available on request.
Software for Detective and Detective-100
Detective/Detective-100 is fully supported by the latest versions
of the highly successful MAESTRO-32 MCA Emulator
as well as the well-known ORTEC Gamma Spectroscopy Packages such as
GammaVision-32 for generalized HPGe spectrum
analysis, PC/FRAM and
MGAHI for Pu and U isotopic ratio analysis
and ISOPlus for in-situ waste assay analysis.
The integral USB connection in the instrument hardware
provides full PC control, real-time live MCA display, fast data transfer
of single and multiple spectra to the PC, and full ORTEC CONNECTIONS
network support.
PLEASE NOTE: MAESTRO-32 (A65-B32) is supplied as
part of packages containing "PKG" in the model number, or it can be
purchased separately at a later date.
|